Working Principle
Purifier HTS automatic nucleic acid extractor adopts magnetic beads separation technology (Figure 2-
4). In this method, a disposable magnetic rod tip comb is loaded on the magnetic rod, and then the magnetic
beads are transferred between the deep hole plates containing specific reagents through the separation of the
magnetic rod and the magnetic rod tip comb, so as to realize the collection, release, transfer and incubation
of the magnetic beads. Different from the traditional nucleic acid extractor, Purifier HTS contains two
heating plates and two standard plates. Only four plates can complete the complete steps of nucleic acid
extraction by magnetic bead method, so as to realize the high-quality extraction of nucleic acid in the
sample, which greatly shortens the extraction time.